Portable Custom Built Device for Thermal Sensitivity Assessment - An Auxiliary Tool to Characterize the Neuropathic Pain following Spinal Cord Injury

Renato Varoto, Fábio Casagrande Hirono, Fernando Ometto Zorzenoni, Ricardo Yoshio Zanetti Kido, William Barcellos, Alberto Cliquet Jr.

2013

Abstract

Neuropathic pain is characterized to arise without stimulation of nociceptors, but due to injury or dysfunction of Peripheral and Central Nervous Systems. It involves altered mechanisms of impulse transmission in somatosensory pathways, causing abnormal sensations. Quantitative sensory testing, by the detection of thermal stimuli, is a method used to characterize and study the neuropathic pain. Therefore, this work describes the development and application of portable custom built device for cutaneous thermal sensitivity assessment in spinal cord injured subjects (SCIS). Using method of levels, the assessment was applied in healthy subjects and SCIS with and without neuropathic pain. The thresholds determined for healthy subjects during thermal sensitivity assessment are consistent and other results provided by clinical trials are according to previous works, demonstrating the device feasibility as an auxiliary tool for neuropathic pain study.

Download


Paper Citation


in Harvard Style

Varoto R., Casagrande Hirono F., Ometto Zorzenoni F., Yoshio Zanetti Kido R., Barcellos W. and Cliquet Jr. A. (2013). Portable Custom Built Device for Thermal Sensitivity Assessment - An Auxiliary Tool to Characterize the Neuropathic Pain following Spinal Cord Injury . In Proceedings of the International Conference on Biomedical Electronics and Devices - Volume 1: BIODEVICES, (BIOSTEC 2013) ISBN 978-989-8565-34-1, pages 28-34. DOI: 10.5220/0004225900280034

in Bibtex Style

@conference{biodevices13,
author={Renato Varoto and Fábio Casagrande Hirono and Fernando Ometto Zorzenoni and Ricardo Yoshio Zanetti Kido and William Barcellos and Alberto Cliquet Jr.},
title={Portable Custom Built Device for Thermal Sensitivity Assessment - An Auxiliary Tool to Characterize the Neuropathic Pain following Spinal Cord Injury},
booktitle={Proceedings of the International Conference on Biomedical Electronics and Devices - Volume 1: BIODEVICES, (BIOSTEC 2013)},
year={2013},
pages={28-34},
publisher={SciTePress},
organization={INSTICC},
doi={10.5220/0004225900280034},
isbn={978-989-8565-34-1},
}


in EndNote Style

TY - CONF
JO - Proceedings of the International Conference on Biomedical Electronics and Devices - Volume 1: BIODEVICES, (BIOSTEC 2013)
TI - Portable Custom Built Device for Thermal Sensitivity Assessment - An Auxiliary Tool to Characterize the Neuropathic Pain following Spinal Cord Injury
SN - 978-989-8565-34-1
AU - Varoto R.
AU - Casagrande Hirono F.
AU - Ometto Zorzenoni F.
AU - Yoshio Zanetti Kido R.
AU - Barcellos W.
AU - Cliquet Jr. A.
PY - 2013
SP - 28
EP - 34
DO - 10.5220/0004225900280034