CAPACITIVE SENSING FOR PULSE RATE MONITORING

R. Niall Tait

2010

Abstract

This paper describes pulse rate monitoring using capacitive electrodes on finger or wrist. The technique uses a single frequency measurement suitable for low cost and low power applications. The system may enable convenient, comfortable, and continuous monitoring. Pulsatile flow resulted in approximately 5-10 fF of capacitance variation, a level easily measured using inexpensive capacitance measurement integrated circuits.

Download


Paper Citation


in Harvard Style

Niall Tait R. (2010). CAPACITIVE SENSING FOR PULSE RATE MONITORING . In Proceedings of the Third International Conference on Biomedical Electronics and Devices - Volume 1: BIODEVICES, (BIOSTEC 2010) ISBN 978-989-674-017-7, pages 215-218. DOI: 10.5220/0002750902150218