VALIDATION OF AN AUTOMATED SEIZURE DETECTION SYSTEM ON HEALTHY BABIES - Histogram-based Energy Normalization for Montage Mismatch Compensation

A. Temko, I. Korotchikova, W. Marnane, G. Lightbody, G. Boylan

2010

Abstract

Seizures in newborn babies are commonly caused by problems such as lack of oxygen, haemorrhage, meningitis, infection and strokes. The aim of an automated neonatal seizure detection system is to assist clinical staff in a neonatal intensive care unit to interpret the EEG. In this work, the automated neonatal seizure detection system is validated on a set of healthy patients and its performance is compared to the performance obtained on sick patients reported previously. The histogram-based energy normalization technique is designed and applied to EEG signals from healthy patients to cope with montage mismatch. The results on healthy babies compares favourably to those obtained on sick babies. Several useful observations are made which were not possible to obtain by testing on sick babies only such as a practically useful range of probabilistic thresholds, minimum detection duration restriction, and an influence of the database statistics on the system performance.

Download


Paper Citation


in Harvard Style

Temko A., Korotchikova I., Marnane W., Lightbody G. and Boylan G. (2010). VALIDATION OF AN AUTOMATED SEIZURE DETECTION SYSTEM ON HEALTHY BABIES - Histogram-based Energy Normalization for Montage Mismatch Compensation . In Proceedings of the Third International Conference on Bio-inspired Systems and Signal Processing - Volume 1: BIOSIGNALS, (BIOSTEC 2010) ISBN 978-989-674-018-4, pages 312-317. DOI: 10.5220/0002750303120317

in Bibtex Style

@conference{biosignals10,
author={A. Temko and I. Korotchikova and W. Marnane and G. Lightbody and G. Boylan},
title={VALIDATION OF AN AUTOMATED SEIZURE DETECTION SYSTEM ON HEALTHY BABIES - Histogram-based Energy Normalization for Montage Mismatch Compensation},
booktitle={Proceedings of the Third International Conference on Bio-inspired Systems and Signal Processing - Volume 1: BIOSIGNALS, (BIOSTEC 2010)},
year={2010},
pages={312-317},
publisher={SciTePress},
organization={INSTICC},
doi={10.5220/0002750303120317},
isbn={978-989-674-018-4},
}


in EndNote Style

TY - CONF
JO - Proceedings of the Third International Conference on Bio-inspired Systems and Signal Processing - Volume 1: BIOSIGNALS, (BIOSTEC 2010)
TI - VALIDATION OF AN AUTOMATED SEIZURE DETECTION SYSTEM ON HEALTHY BABIES - Histogram-based Energy Normalization for Montage Mismatch Compensation
SN - 978-989-674-018-4
AU - Temko A.
AU - Korotchikova I.
AU - Marnane W.
AU - Lightbody G.
AU - Boylan G.
PY - 2010
SP - 312
EP - 317
DO - 10.5220/0002750303120317